GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-17
8-2-10
Control Console Bottom Cover (Upper Section) Replacement Procedure
8-2-10-1
Tools
Use the appropriate Phillips-type screw drivers, as indicated in the replacement procedures of the upper
section of the control console’s bottom cover.
8-2-10-2
Preparation
Shut down the Vivid 3N Pro/Expert
ultrasound unit, as described in
Chapter 3 -
Installation
.
8-2-10-3
Control Console Bottom Cover (Upper Section) Removal Procedure
1) Raise the control console to its
maximum
height.
2) Remove any peripherals.
3) Unplug the AC cable and VGA cable from the rear of the control console, as shown below:
Figure 8-18 AC and VGA Cables at Rear of Control Console
4) Remove the two screws that fasten the rear section of the gas spring cover, and remove this
section.
5) Remove the three screws that fasten the front section of the gas spring cover, and allow this section
to sink down into the top cover collar.
6) Remove the two screws located on the white painted metal shelf inside the compartment that is
located on the left side of the control console, and remove this shelf.
AC Cable
VGA Cable
Содержание Vivid 3N Pro Series
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii...
Страница 597: ......