Section 5-7 - Vivid 3N Power Distribution
5-7-3-3
Input AC Voltage Configuration
•
The Internal Vivid 3N
voltage is 220 V AC.
•
Back End and Front End Crates operate on 220 V AC.
•
The Monitor and Peripherals operate on local voltage.
The AC input voltage is factory
preset in accordance with specific requirements, as indicated on the
label near the AC input connector. The AC transformer’s primary and secondary voltages are set up by
using the appropriate plugs connected in Jumper1 and Jumper2 on the AC Distribution Box, according
to the following table:
NOTE:
The AC input voltage power cord and its plug must meet the local standard requirements.
5-7-4
Front End DC Power Distribution
Table 5-6
Primary and Secondary AC Voltage
Jumper 1
PRIMARY
Voltage Range
100-120V
220-240V
Plug P/N
2253054
2253056
Jumper 2
SECONDARY
Voltage
100V
120V
220-230V
230-240V
Plug P/N
2253057
2253058
2253059-2
2253060-2
Figure 5-28
DC Distribution: Front End Crate ( RFI Configuration)