Section 3-9 - Connectivity Setup
3-9-7
Image Vault Server
If
Image Vault Server
is required, make sure that the
Image Vault Server
checkbox is
selected
, as
shown in
Figure 3-47
below, then proceed as follows:
a.) Enter
ECHOPAC7-OOOXXX-DB
in the
Database Name
field.
b.) Under
Store Info,
make sure that the
Hard Drive
checkbox is selected.
c.) Enter
\\ECHOPAC7-000XXX\Archive
in the
Placement of Image Files
field.
d.)
Select
the
Allow Multiframe
and/or
Allow Raw Data
checkboxes, as required.
(It is recommended to leave both checkboxes activated, as shown in
Figure 3-47
below).
e.) Under the
Image Settings
field, select the required settings.
Figure 3-47 Services Tab - Image Vault Server Expanded