Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-6 Pencil Probe Test
(cont’d)
7) Check the test result message and proceed as follows:
•
If one or more relays were identified as bad, trackball to
Save Report
and press
Select
, and
contact a GE Medical Systems service representative.
•
If no relays were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to the channel in the Display area and press
Select
.
•
Trackball to the
Mark bad channel
checkbox and press
Select
. Repeat as required to
mark additional channels.
•
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
•
If no errors were identified, proceed with the next test.
•
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 -
Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
8) Trackball to the
Exit
button and press
Select
. If required, repeat this test with a different pulser.