Chapter 7 Diagnostics/Troubleshooting
7-89
7-5-2-1
Service Log In
1) From the System Configuration window, make sure that
iLinq
(the Common Service platform) is
enabled under the
Technical Support
tab, as shown in
Figure 7-59
below.
Figure 7-59 Technical Support - System Configuration
2) Trackball to
OK
and press
Select
. The System Configuration window closes.
NOTE:
If iLinq was not found to be enabled in
step 1
, it is necessary to re-boot the system before
proceeding.
3) Trackball to the wrench icon
in the status bar at the bottom of the scanning screen and press
Select
.
The
Profile Manager
window is displayed, as shown in
Figure 7-60 "Profile Manager Window" on
page 7-90
.
RFT-configured systems
RFI-configured systems