10-26
Section 10-7 - Electrical Safety Tests
10-7-8-1
Data Sheet for ECG Current Leakage
The test passes when all readings measure less than the values shown in the tables below. Record all
data on the
Vivid 3N Preventative Maintenance Inspection Certificate
(see page 10-31).
Table 10-18 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
AC Power
Source
Maximum
Allowance Limit
GROUND
OPEN
GROUND
CLOSED
Patient Lead-to-Ground Current Leakage Test
and
Patient Lead-to-Lead Current Leakage Test
115V
10uA
10uA
220/240V
500uA
10uA
Table 10-19 Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power
Source
Maximum
Allowance Limit
Patient Lead Isolation Current Test
115V
20uA
220/240V
5mA
Table 10-20 Typical Data Sheet for ECG Lead-to-Lead Test
ECG
Power
Tester
Polarity
Switch
Tester
Ground
Switch
Tester Lead Selector
RL
(N)
RA
(R)
LA
(L)
LL
C
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN