GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
3-44
Section 3-8 - System Configuration
3-8-9
VCR/ECG Tab
1) Press
Config
on the alphanumeric keyboard. The
System Configuration
window is displayed.
2) Trackball to the
VCR/ECG
tab and press
Select
. The
VCR/ECG
tab is displayed, as shown below:
Figure 3-27 VCR/ECG Tab
3) In the
VCR
Options
area, complete the following, as required:
•
Select the video system from the
Video Mode
dropdown list, for example, PAL or NTSC.
•
Select the
SmartVCR
option to enable Measurement and Analysis measurements and
calculations on recorded VCR images.
•
Select the
Patient Name on second line
option to display the patient’s name on the second
line. For use when an external monitor omits picture borders.
4) If done, trackball to the
OK
button and press
Select
.
Содержание Vivid 3N Pro Series