GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
4-6
Section 4-3 - Functional Check
4-3-3
Mechanical Functions
4-3-3-1
Console Movement Test
4-3-3-2
Brakes Function Test
Table 4-11 Console Movement Test
Step
Task
Expected Result(s)
1
Pull the front handle located under the control console
and lift the console to its maximum height.
Ensure that you do not apply too much force to move
the console and that the console movement is smooth.
2
Pull the front handle located under the control console
and push the console to its minimum height.
Ensure that you do not apply too much force to move
the console and that the console movement is smooth.
Table 4-12 Brakes Function Test
Step
Task
Expected Result(s)
1
Push the brake pedal to its
center
position. Push and pull
the unit right, left, backwards and forwards.
Ensure that the wheels move freely in all directions.
Check the wheels for wear and tear, and replace if
necessary.
2
Push the brake pedal to its
left
position (parking). Push
and pull the unit right, left, backwards and forwards.
Ensure that the wheels are locked and there is no
movement in any direction.
3
Push the brake pedal to its
right
position (arrows). Push
and pull the unit right, left, backwards and forwards.
Ensure that the front wheels are locked for right and left
movement
only
, but can move freely backwards and
forwards.