GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-65
8-5-3-5
BEP2 - Front Cover Removal Procedure
1) Remove the two screws that secure the BEP front cover in position - see
Figure 8-60 "Front Cover
- BEP2" on page 8-65
.
2) Press on the two latches on the side to release the cover.
3.) Pull the cover forward (see
Figure 8-61 "Front Cover Removal - BEP2" on page 8-66
) and remove it.
Figure 8-60 Front Cover - BEP2