GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-56
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9-1 Accessing the MUX Diagnostic Test Options
(cont’d)
6) Check the test result message and proceed as follows:
•
If one or more channels were identified as bad, trackball to the
Save Report
button and press
Select
, then contact a GE Medical Systems Service Representative.
•
If no channels were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to each of the channels in the
Display
area and press select.
•
Trackball to the
Mark bad channel
checkbox and press
Select
.
•
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
•
If no errors were identified, proceed with the next test.
•
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 -
Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
7) Trackball to the
Exit
button and press
Select
. If required, repeat this test with a different pulser.
Содержание Vivid 3N Pro Series
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii...
Страница 597: ......