Chapter 4 Functional Checks
4-7
4-3-4
Back End Processor Tests
4-3-4-1
Internal ECG Test
4-3-4-2
Phono Test (optional)
Table 4-13 Internal ECG Test
Step
Task
Expected Result(s)
1
Connect the internal ECG input to an ECG simulator or
suitable substitute, using the cable set and electrode
pads.
2
Perform the ECG diagnostic test, as described in
7-4-3
"ECG/Phono Diagnostic Test" on page 7-74
.
Verify that a single clear QRS signal is displayed.
Table 4-14 Phono Test
Step
Task
Expected Result(s)
1
Connect the microphone to the microphone input on the
ECG Board. Note that the special Medical Microphone is
required.
2
Perform the Phono diagnostic test, as described in
7-4-3 "ECG/Phono Diagnostic Test" on page 7-74
Use your finger to click (tap) the microphone.
Verify that the waveform on the screen corresponds to
the microphone clicks.