GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
8-120
Section 8-8 - Peripherals
8-8-1-4-1
Setting the Switches on the Paper Tray
1.) Press the OPEN button in the front of the B/W Video Printer.
2.) Ensure that the switches on the Paper Tray are set as shown in
Figure 8-103
below.
3.) Load the Paper Roll as described in the Printer’s Operating Instructions.
8-8-1-5
B&W Printer Removal Procedure
1.) Remove the Connectors Panels Cover as described in
8-2-5-3 "Connector Panels Cover Removal
Procedure" on page 8-8
.
2.) If optional shelf is present, refer to
8-8-4-5 "Color Video Printer Removal - Procedure 1" on page 8-
139
.
3.) Release the B/W printer by undoing the fastening belt secured around it (and around the VCR, if
present).
4.) Disconnect the Remote cable, the B/W Video IN cable, and Power cable from rear of the printer
(refer to
Figure 8-102 "B/W Video Printer Connection Cables" on page 8-119
).
5.) Remove the B/W Video Printer from the
right
side of the Vivid 3N ultrasound unit (taking it out from
below the control console and from beside the B/W printer, if present) - refer to
Figure 8-100 "B/W
Video Printer Installed in the Vivid 3N |(Configuration for UP2950 and 2800 Printers)" on page 8-
118
.
Figure 8-103 B/W Video Printer, Slide Switches Location and Settings on the Paper Tray
Slide Switch
Settings on the
Paper Tray
OPEN Tray button