Chapter 7 Diagnostics/Troubleshooting
7-67
7-3-9-7-2 Temp Test
The Temp test displays the temperature at the Front End crate, as well as the temperature of the AC
distribution box.
1) Access the
MUX
tab of the
FB Diagnostics
dialog box, as described in
7-3-9-1 "Accessing the MUX
Diagnostic Test Options" on page 7-50
.
2) Trackball to the
Advanced board tests
button and press
Select
. The
Advanced System Tests
options are displayed, as shown in
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
61
.
3) Trackball to the
Monitor
button and press
Select
. The
View Power
options are displayed, as shown
in
Figure 7-37 "View Power Options" on page 7-63
.
4) Trackball to the
Temperature
button and press
Select
. The temperatures are displayed, as shown
below:
Figure 7-41 Temperatures
5) Trackball to the
Show
button and press
Select
to display the temperatures.
NOTE:
For repeat readings, select
Loop every
and set the reading time cycle.