GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 2 Pre-Installation
2-9
2-3-3-1
Recommended Ultrasound Room Layout
Figure 2-3 Minimal Floor Plan 2.5m x 3m (8.2ft x 9.84 ft)
2-3-4
Networking Pre-Installation Requirements
2-3-4-1
Stand-alone Unit (without Network Connection)
None.
2-3-4-2
Unit Connected to Hospital’s Network
Supported networks:
•
100/10 Mbit/sec
2-3-4-3
Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
VIVID 3
36 IN.
(92 CM)
Scale: Each square = 1 sq ft (144 sq ins)
Dedicated Power Outlets
Hospital Network
Dedicated Analog Telephone Line
for Connection to InSite
GE Cabinet for
Software and Manuals
(optional)
Содержание Vivid 3N Pro Series
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii...
Страница 597: ......