Chapter 7 Diagnostics/Troubleshooting
7-47
7-3-8-7
Port Test
The Port test enables you to check the general logic control and ports logic.
NOTE:
You must reboot the system after performing the Port test.
1) Disconnect all probes.
2) Access the
F.B.
tab of the
FE Diagnostics
dialog box, as described in
7-3-8-1 "Accessing the Front
Board Assembly Options" on page 7-27
.
3) Trackball to the
Port Test
button and press
Select
. The port test is performed and the results are
displayed in the
Comments
area.
4) Check the test results and proceed as follows:
•
If no errors were identified, proceed with the next test.
•
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 -
Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
5) Reboot the system.