Chapter 8 Replacement Procedures
8-91
8-5-14-4
Hard Disk Installation Procedure
Note:
Before installing a new hard disk, make sure the jumper settings are as shown below:
1) Place the new hard disk in position in the hard disk module holder.
2.) Reconnect all the cables at the rear of the hard disk.
3) Fasten each of the four screws (two on each side), as shown in
Figure 8-77 "Hard Disk Removal 2"
on page 8-90
.
4) Carefully slide the module holder into position in the unit.
5) Return and fasten the screw that holds the hard disk module in position (as shown in
Figure 8-74
"PC-VIC" on page 8-87
6) Install the BEP side cover, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-4 "BEP2 -
Side Cover Installation Procedure" on page 8-64
.
7) Return the Vivid 3N left cover, as described in
8-2-2-4 "Side Covers Installation Procedure" on page
8-4
.
Note:
When a new hard disk arrives from the manufacturer, usually it is partitioned and ready for
software installation.
Refer to
Section 8-7 "Software Loading" on page 8-115
for software installation instructions.