Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-4 Receive Test
(cont’d)
9.) Check the test result message and proceed as follows:
•
If one or more channels were identified as bad, trackball to the
Save Report
button and press
Select
,
then contact a GE Medical Systems Service Representative.
•
If no channels were identified as bad, mark the channels suspected as being bad as follows:
-
Trackball to each of the channels in the
Display
area and press select.
-
Trackball to the
Mark bad channel
checkbox and press
Select
.
-
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
•
If no errors were identified, proceed with the next test.
•
If no errors were identified but the problem persists, contact your local GE Medical Systems field
engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration
section, on page 6-14, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI bridge
cables are connected and functioning properly, as described in
Chapter 5 - Components and Function
(Theory)
. If the problem persists, contact your local GE Medical Systems field engineer for
assistance.
10.)Trackball to the
Exit
button and press
Select
. If required, repeat this test with a different pulser.