Chapter 7 Diagnostics/Troubleshooting
7-85
7-4-7-2
Check Disk/Bus
1) Access the
HW Diagnostics
menu, as described in
7-4-1 "Accessing the Back End Diagnostic
Options" on page 7-71
.
2) Trackball to the
Computer Testing
button and press
Select
. The options shown in
Figure 7-54
"Benchmark Testing Options" on page 7-83
are displayed.
3) Trackball to the
Check Disc/Bus
button and press
Select
. The
Disk/Bus Bench
window is
displayed as shown in
Figure 7-56 "Disk/Bus Bench Window" on page 7-85
.
Figure 7-56 Disk/Bus Bench Window
Disk Bench
: In the information displayed, make sure that the speed values displayed are greater than
23MB/sec. This test checks the speed of information transfer between the hard disk and IDE BUS.
Bus Bench
: In the information displayed, make sure that the speed values displayed are greater than
86MB/sec. This test checks the speed of information transfer between the memory and the BUS.