GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
10-10
Section 10-5 - System Periodic Maintenance
10-5-4
Cleaning
10-5-4-1
General Cleaning
10-5-4-2
Air Filter Cleaning
Regular cleaning of the Vivid 3N scanner’s air filter will ensure that the filter does not become clogged
- this would otherwise cause the system to overheat and reduce system performance and reliability.
NOTE:
It is recommended that the air filter cleaning procedure is performed once every 3 months (quarterly).
However, the required frequency for air filter cleaning will vary in accordance with environmental
conditions.
Table 10-8 General Cleaning
Step
Item
Description
1
External
Panels
On a weekly basis, moisten a soft, non-abrasive folded cloth with a mild, general purpose,
non-abrasive soap and water solution or general purpose disinfectant. Wipe down the top,
front, back and both sides of the unit
.
Do not spray any liquid directly onto the unit!
2
Monitor
On a weekly basis, apply glass cleaner to a soft, non-abrasive folded cloth. Gently wipe the
monitor face.
3
Console
Frequent and diligent cleaning of the Vivid 3N Pro/Expert ultrasound unit reduces the risk of
spreading infection from person to person, and also helps to maintain a clean working
environment.
Prior to cleaning, turn OFF the power to the system.
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
4
Probe Holder
Clean the probe holders with warm water and a damp cloth to remove all traces of gel
(soaking may be required to remove excess gel).
Table 10-9 Air Filter Cleaning
Step
Item
Description
1
Air Filter
Remove the air filter located at the front of the scanner, above the probe connector sockets.
Clean the air filter as follows: shake it in an area away from the system, wash with a mild
soapy solution, rinse, and air dry.
2
Air Filter
Install the clean air filter.
Refer to
Chapter 8 -
for air filter installation instructions.