Section 8-5 - Back End Parts Replacement
8-5-8
PC2IP Board Replacement Procedure
8-5-8-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the PC2IP board
replacement procedure.
8-5-8-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-5-8-3
PC2IP Board Removal Procedure
1) Remove the Vivid 3N left and right side covers, and rear cover - see the
8-2-4-3 "Rear Cover
Removal Procedure" on page 8-7
and the
8-2-2-4 "Side Covers Installation Procedure" on page 8-4
.
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP side covers, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-3 "BEP2 -
Side Cover Removal Procedure" on page 8-64
.
4) If applicable, remove the octopus card holder as described in
8-5-3-7 "BEP2 - Octopus Card Holder
Removal Procedure" on page 8-67
.
5) Disconnect the PCI cable that is connected to the BEP.
Note:
Counting from the top, the PC2IP board is positioned in the fourth slot.
6) Release the securing screw located on the left of the board bracket.
7) Gently remove the PC2IP board from the BEP.
8-5-8-4
PC2IP Board Installation Procedure
1) Gently insert the PC2IP board into its slot.
Note:
Counting from the top, the PC2IP board is positioned in the fourth slot.
2) Fasten the board with the screw on the left of the bracket.
3) Reconnect the PCI cable that is connected to the BEP.
4.) If applicable, install the octopus card holder as described in
8-5-3-8 "BEP2 - Octopus Card Holder
Installation Procedure" on page 8-67
.
5) Reconnect the power cable to the rear of the BEP.
6.) Install the BEP
side covers, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-4 "BEP2 -
Side Cover Installation Procedure" on page 8-64
.
7) Return the Vivid 3N rear, left and right covers - see the
8-2-2-4 "Side Covers Installation Procedure"
on page 8-4
and
8-2-4 "Rear Cover Replacement Procedures" on page 8-7
.
Note:
Perform a complete diagnostic check of the system, as described in
Chapter 7 -
Diagnostics/Troubleshooting
.