GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-92
Section 7-5 - Common Service Interface
7-5-2-2
Access and Security
The service interface has various different security and user levels. Users are only granted access to
the tools they are authorized to use, as described in the following table:
All access to the service interface is via the Network port, except for iLinq. A modem on the scanner is
specifically intended for iLinq use. Every access request, whether successful or not is logged in a
service access log that can be viewed by authorized users.
7-5-3
Error Logs Page
When the
Error Logs
page is selected (as shown in
Figure 7-63
below), the log viewing options described
in the following sections, are available:
•
Log Viewer
section, on page 7-93.
•
Log Browser
section, on page 7-98.
Figure 7-63 Error Logs Page
Table 7-1: Access and Security
User Level
Access Authorization
Operator
Normally used in-house and protected with a special scanner configuration that is
not allowed to leave the facility.
Administrator
Use the OLC access method provided by iLinq.
External Services
Access is protected with a physical media key (typically a CD-ROM) and a
password. The contents of the media key determine the level of access and the
period of time for which access is granted.
GE Service
Requires a network connection, a physical media key and knowledge of the
password.