GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
5-22
Section 5-4 - Front End
5-4-8
Front End Controller Board (FEC)(RFT)
The Front End Controller (FEC) Board controls other boards in the Front End Crate. The control is done
through a synchronous and bidirectional Bus called the Front End Bus. The following tasks are
performed by the FEC Board:
•
System master clock generation (including clock to IP).
•
Transmitter control.
•
ATCG control.
•
Receive focusing control.
•
Analog test signal control.
•
Probe sensing.
•
Transmit voltage control.
•
Control of Front Board switching.
•
Temperature sensing for the TEE probes.
•
Memory refresh control for the BF Board.
•
Provide 40 MHz clock to IP.
Figure 5-1: Front End Controller Block Diagram
FE Bus P2
FB, TX, RFT Trig Ps-A25
FB P-B35
FB, Probe CTRL P3
FB, ATGC P2-B37+
P2-D37-
All Boards P3-D25
P3-D26
Front End Bus
Interface
AGTC
Generator
Test Signal
Generator
XDCCTRL
Interface
EI Probe
ID/Select
Revision Detect
System Clock
Generator
VMA Bus
Interface
HV Control
HV Readback
CPU
Memory
P3 FB
P3 FB
VME Bus
P IP