GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-68
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9-7-3 Show History
The
Show History
option in the
View Power
dialog box displays selected test results by
year
,
month
and
day
.
1) Access the
MUX
tab of the
FB Diagnostics
dialog box, as described in
7-3-9-1 "Accessing the MUX
Diagnostic Test Options" on page 7-50
.
2) Trackball to the
Advanced board tests
button and press
Select
. The
Advanced System Tests
options are displayed, as shown in
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
61
.
3) Trackball to the
Monitor
button and press
Select
. The
View Power
options are displayed, as shown
in
Figure 7-37 "View Power Options" on page 7-63
.
4) Trackball to the
View History
button and press
Select
. The following options are displayed.
Figure 7-42 View History Options
5) Select the test and date that you want to view from the appropriate drop-down list, then trackball to
the
View
button and press
Select
. The relevant graphs are displayed showing activity over time in
values of
Temperature
,
%
, or
Voltage
.
Содержание Vivid 3N Pro Series