Chapter 8 Replacement Procedures
8-67
8-5-3-7
BEP2 - Octopus Card Holder Removal Procedure
1) Remove the three screws that secure the Octopus Card Holder in position as shown in
Figure 8-62
below.
2.) Remove the Octopus Card Holder.
8-5-3-8
BEP2 - Octopus Card Holder Installation Procedure
1) Place the Octopus Card Holder in position on the BEP part, holding it firmly in place.
2) Return the three screws and tighten them to hold the Octopus Card Holder securely in position, as
shown in
Figure 8-62
, above.
Figure 8-62 Octopus Card Holder - BEP1