Remote Commands
R&S
®
FPL1000
795
User Manual 1178.3370.02 ─ 11
[SENSe:]CORRection:TRANsducer:GENerate
<Name>
This command uses the normalized measurement data to generate a transducer factor
with up to 1001 points. The trace data is converted to a transducer with unit dB and
stored in a file with the specified name and the suffix
.trd
under
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\trd
. The frequency points are allo-
cated in equidistant steps between start and stop frequency.
The generated transducer factor can be further adapted using the commands descri-
bed in
Chapter 10.10.3, "Working with Transducers"
Parameters:
<Name>
'<name>'
Example:
CORR:TRAN:GEN 'MyGenerator'
Creates the transducer file
C:\r_s\instr\trd\MyGenerator.trd
.
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
<Value>
This command defines the reference value assigned to the reference position in the
specified window. Separate reference values are maintained for the various displays.
Suffix:
<n>
.
<w>
subwindow
<t>
irrelevant
Parameters:
<Value>
Default unit: DB
Example:
DISP:TRAC:Y:RVAL 0
Sets the value assigned to the reference position to 0 Hz
Manual operation:
See
Power Sweep Commands
The following commands are required to configure a power sweep using an internal
tracking generator.
For an example of performing a power sweep remotely, see
gramming Example: Measuring Compression Points"
Useful commands for power sweeps described elsewhere:
●
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
●
SOURce<si>:INTernal:FREQuency:OFFSet
●
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
●
Setting Basic Spectrum RF Measurement Parameters