The Spectrum Application (RF Measurements)
R&S
®
FPL1000
264
User Manual 1178.3370.02 ─ 11
Remote command:
CALCulate<n>:DELTamarker<m>:MREFerence
Linking to Another Marker
Links the current marker to the marker selected from the list of active markers. If the x-
axis value of the initial marker is changed, the linked marker follows to the same posi-
tion on the x-axis. Linking is off by default.
Using this function you can set two markers on different traces to measure the differ-
ence (e.g. between a max hold trace and a min hold trace or between a measurement
and a reference trace).
Remote command:
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<m>:LINK
Assigning the Marker to a Trace
The "Trace" setting assigns the selected marker to an active trace. The trace deter-
mines which value the marker shows at the marker position. If the marker was previ-
ously assigned to a different trace, the marker remains on the previous frequency or
time, but indicates the value of the new trace.
If a trace is turned off, the assigned markers and marker functions are also deactiva-
ted.
Remote command:
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency.
This setting is only available if the EMI (R&S
FPL1-K54) measurement option is instal-
led. CISPR detectors are only available after selecting a CISPR RBW or filter (see
For details, see
Note:
The trace detector configured in the trace settings is used for the initial peak
search only, see
Chapter 6.10.1.2, "Trace Settings"
"Off"
No final test is performed.
"PositivePeak"
Determines the maximum signal level that was detected during the
specified dwell time.
"Average"
Determines the average signal level of the samples that were collec-
ted during the specified dwell time.
"Quasi-Peak"
Determines the maximum signal level weighted to CISPR
16-1-1 that
was detected during the dwell time.
The "Quasi-Peak" detector is only available for the CISPR filter, and
not for an RBW of 1
MHz.
Measurements and Results