The Spectrum Application (RF Measurements)
R&S
®
FPL1000
266
User Manual 1178.3370.02 ─ 11
The detector to be used for the final EMI test can be defined differently for each fre-
quency, thus the detector is configured in the EMI marker settings, see
...............................................................................................267
...............................................................................................268
Filter Type
Defines the filter type.
The following filter types are available:
●
Normal (3dB)
●
Channel
●
CISPR (6 dB) - requires EMI (R&S
FPL1-K54) option
●
MIL Std (6 dB) - requires EMI (R&S
FPL1-K54) option
For more information, see
Chapter 6.6.1.6, "Which Data May Pass: Filter Types"
Note:
The EMI-specific filter types are available if the EMI (R&S
FPL1-K54) measure-
ment option is installed, even if EMI measurement is not active. If you select a CISPR
trace detector, the filter type is automatically also set to CISPR. An active CISPR filter
Measurements and Results