Remote Commands
R&S
®
FPL1000
735
User Manual 1178.3370.02 ─ 11
Parameters:
<Frequency>
Defines the pulse frequency.
Range:
0 to Fmax
Default unit: Hz
<RBW>
Defines the resolution bandwidth.
Default unit: HZ
<MeasTime>
Defines the measurement time.
Range:
1 µs to 30 s
Default unit: S
<TriggerSource>
Selects a trigger source.
<TriggerLevel>
Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S
FPL1000 uses a fix TTL
level.
Range:
0 to 100
Default unit: PCT
<TriggerOffset>
Defines the trigger delay.
Range:
0 s to 30 s
*RST:
0 s
Default unit: s
<Detector>
Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
<OfPulses>
Defines the number of pulses included in the measurement.
Range:
1 to 32001
Return values:
<PowerLevel>
Pulse power level.
The result is a list of floating point values separated by commas.
The unit is dBm.
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Return values:
<PulsePower>
Lowest power level of the pulse power measurement.
The unit is dBm.
Configuring and Performing Measurements