Remote Commands
R&S
®
FPL1000
965
User Manual 1178.3370.02 ─ 11
............................................................................................. 967
............................................................................................ 968
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
This command sets the GPIB address of the R&S
FPL1000.
Parameters:
<Address>
Range:
0 to 30
*RST:
(no influence on this parameter, factory default 20)
Example:
SYST:COMM:GPIB:ADDR 18
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
This command selects the GPIB receive terminator.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator>
LFEOI | EOI
LFEOI
According to the standard, the terminator in ASCII is <LF>
and/or <EOI>.
EOI
For binary data transfers (e.g. trace data) from the control com-
puter to the instrument, the binary code used for <LF> might be
included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be
avoided by using only the receive terminator
EOI
.
*RST:
LFEOI
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation:
See
SYSTem:DISPlay:LOCK
<State>
Defines whether the "Display Update" function remains available in remote operation
or not.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
The function remains available.
Configuring the R&S
FPL1000