Remote Commands
R&S
®
FPL1000
775
User Manual 1178.3370.02 ─ 11
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
"Incompleted sequential commands - blocked remote
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the "Trigger Input" connector.
Trigger signal from the "Trigger In" connector.
IFPower
Second intermediate frequency
TIME
Time interval
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ns to 5000 s
*RST:
1.0 s
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters