![Ronde & Schwarz 1304.0004K03 User Manual Download Page 121](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987121.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
121
User Manual 1178.3370.02 ─ 11
The settings for a power sweep measurement are described in
About the Power Sweep Compression Point Measurement
................................. 121
Power Sweep Compression Point Results
............................................................123
Power Sweep Compression Point Settings
...........................................................124
How to Determine Compression Points
................................................................ 126
6.2.3.1
About the Power Sweep Compression Point Measurement
The compression point refers to the power level at which a power-converting device
deviates from a linear input to output relation. The compression point of an amplifier
defines the upper limit of its dynamic range. Using an optional internal generator, the
R&S
FPL1000 can perform a power sweep to determine the compression point of a
device under test.
In a power sweep compression point measurement, you use the internal generator sig-
nal as a reference input signal for your power-converting DUT. Then you compare the
DUT input to its output, which is measured at the RF input of the analyzer.
Measurements and Results