User Manual 1178.3370.02 ─ 11
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
<w>
subwindow
<t>
irrelevant
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
FPL1000 adjusts the scaling of the y-axis accordingly.
For measurements with the optional tracking generator, the command defines the posi-
tion of the reference line.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Position>
0 PCT corresponds to the lower display border, 100 percent cor-
responds to the upper display border.
*RST:
frequency display: 90 PCT; time display: 50 PCT;
AF spectrum display (K7): 100 PCT;
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation:
See
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
<w>
subwindow
Setting Basic Spectrum RF Measurement Parameters