![Ronde & Schwarz 1304.0004K03 User Manual Download Page 269](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987269.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
269
User Manual 1178.3370.02 ─ 11
Res BW CISPR
Automatically sets the measurement bandwidth for commercial EMC standards
according to CISPR.
For more information, see
"Resolution Bandwidth and Filter Types"
Remote command:
Filter type:
[SENSe:]BANDwidth[:RESolution]:TYPE
Filter bandwidth:
[SENSe:]BANDwidth[:RESolution]
Res BW MIL
Automatically sets the measurement bandwidth for military EMC standards.
For more information, see
"Resolution Bandwidth and Filter Types"
Remote command:
Filter type:
[SENSe:]BANDwidth[:RESolution]:TYPE
Filter bandwidth:
[SENSe:]BANDwidth[:RESolution]
LISN Control Settings
Access
: [MEAS CONFIG] > "LISN Config"
For measurements with power lines, the following settings are available for the
R&S
FPL1000 to control which phase of the LISN is to be tested (e.g. for EMI mea-
surements). LISN control requires the EMI measurement (R&S
FPL1-K54) option.
It also requires the optional additional interfaces (R&S
FPL1-B5).
Measurements and Results