![Ronde & Schwarz 1304.0004K03 User Manual Download Page 874](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987874.webp)
Remote Commands
R&S
®
FPL1000
874
User Manual 1178.3370.02 ─ 11
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
<State>
This command turns delta markers for band power measurements on and off.
If neccessary, the command also turns on a reference marker.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
"Band Power Measurement State"
See
"Switching All Band Power Measurements Off"
10.8.8.12
n dB Down Marker
The following commands control the n dB down markers.
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
............................................................ 874
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
........................................875
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:QFACtor?
............................................. 875
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?
............................................... 876
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe
.................................................. 876
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
..................................................877
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
<Distance>
This command defines the distance of the n dB down markers to the reference marker.
Suffix:
<n>
.
<m>
Parameters:
<Distance>
Distance of the temporary markers to the reference marker in
dB.
For a positive offset, the markers T1 and T2 are placed
below
the active reference point.
For a negative offset (for example for notch filter measure-
ments), the markers T1 and T2 are placed
above
the active ref-
erence point.
*RST:
6dB
Default unit: DB
Setting Basic Spectrum RF Measurement Parameters