Remote Commands
R&S
®
FPL1000
813
User Manual 1178.3370.02 ─ 11
CALCulate<n>:SGRam:CLEar[:IMMediate]
CALCulate<n>:SPECtrogram:CLEar[:IMMediate]
This command resets the spectrogram and clears the history buffer.
Suffix:
<n>
.
Example:
//Reset the result display and clear the memory
CALC:SGR:CLE
Manual operation:
See
CALCulate<n>:SGRam:CONTinuous
<State>
CALCulate<n>:SPECtrogram:CONTinuous
<State>
This command determines whether the results of the last measurement are deleted
before starting a new measurement in single sweep mode.
This setting applies to all spectrograms in the channel setup.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
INIT:CONT OFF
Selects single sweep mode.
INIT;*WAI
Starts the sweep and waits for the end of the sweep.
CALC:SGR:CONT ON
Repeats the single sweep measurement without deleting the
results of the last measurement.
Manual operation:
See
See
CALCulate<n>:SGRam:FRAMe:COUNt
<Frames>
CALCulate<n>:SPECtrogram:FRAMe:COUNt
<Frames>
This command defines the number of frames to be recorded in a single sweep.
This value applies to all spectrograms in the channel setup.
Suffix:
<n>
.