![Ronde & Schwarz 1304.0004K03 User Manual Download Page 762](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987762.webp)
Remote Commands
R&S
®
FPL1000
762
User Manual 1178.3370.02 ─ 11
Remote commands exclusive to configuring sweeps:
.................................................................................................762
........................................................................................... 762
.......................................................................................763
.............................................................................................763
[SENSe:]SWEep[:WINDow<n>]:POINts
............................................................................764
......................................................................................... 765
.........................................................................................765
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In continuous sweep mode, the application calculates the moving average over the
average count.
In single sweep mode, the application stops the measurement and calculates the aver-
age after the average count has been reached.
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S
FPL1000 per-
forms one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:DURation?
<Time>
This command provides an estimation of the total time required to capture the data and
process it. This time span may be considerably longer than the actual sweep time (see
Tip:
To determine the necessary timeout for data capturing in a remote control pro-
gram, double the estimated time and add 1 second.
Return values:
<Time>
Setting Basic Spectrum RF Measurement Parameters