Remote Commands
R&S
®
FPL1000
710
User Manual 1178.3370.02 ─ 11
10.6.11.1
Activating the Measurement
The following command activates harmonic distortion measurement.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
.............................................. 710
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
<State>
This command turns the harmonic distortion measurement on and off.
Note the following:
●
If you perform the measurement in the frequency domain, the search range for the
frequency of the first harmonic, whose power is determined, is defined by the last
span.
●
If you perform the measurement in the time domain, the current center frequency is
used as the frequency of the first harmonic. Thus, the frequency search is
bypassed. The first harmonic frequency is set by a specific center frequency in
zero span before the harmonic measurement is started.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
10.6.11.2
Configuring the Measurement
The following commands control the harmonic distortion measurement.
Useful commands for harmonic distortion measurements described elsewhere
●
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
●
Remote commands exclusive to harmonic distortion measurements
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
............................... 710
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics
...................................... 711
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet
.............................................. 711
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
<State>
This command selects the resolution bandwidth of the harmonic in respect to the band-
width of the first harmonic.
Configuring and Performing Measurements