![Ronde & Schwarz 1304.0004K03 User Manual Download Page 682](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987682.webp)
Remote Commands
R&S
®
FPL1000
682
User Manual 1178.3370.02 ─ 11
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO
<State>
This command turns automatic selection of the input attenuation for a spurious emis-
sion measurement range on and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
LIST:RANG2:INP:ATT:AUTO ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
LIST:RANG2:INP:GAIN:STAT ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:INP:GAIN 15
[SENSe:]LIST:RANGe<ri>:LIMit:STARt
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Configuring and Performing Measurements