Remote Commands
R&S
®
FPL1000
622
User Manual 1178.3370.02 ─ 11
10.6.3.1
Configuring the Compression Point Measurement
The following commands are required to configure compression point measurements
using an internal tracking generator.
For details see
Chapter 6.2.3, "Power Sweep Measurements"
For a programming example using remote commands, see
gramming Example: Measuring Compression Points"
Useful commands for compression point measurements described elsewhere:
●
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
Remote commands exclusive to compression point measurements:
CONFigure:CMEasurement:REFerence:VALue
.................................................................622
CONFigure:CMEasurement:REFerence[:MODE]
...............................................................622
CONFigure:CMEasurement:RESult:CVALue
.....................................................................623
CONFigure:CMEasurement:RESult:P1DB[:STAT]
..............................................................623
CONFigure:CMEasurement:RESult:P3DB[:STAT]
..............................................................624
CONFigure:CMEasurement:RESult:PNDB[:STAT]
............................................................. 624
CONFigure:CMEasurement[:STATe]
.................................................................................624
FETCh:CMEasurement:P1DB[:RESult]?
...........................................................................625
FETCh:CMEasurement:P3DB[:RESult]?
...........................................................................625
FETCh:CMEasurement:PNDB[:RESult]?
.......................................................................... 625
CONFigure:CMEasurement:REFerence:VALue
<ReferenceValue>
Defines the reference value used to calculate the compression point for
MANual
.
Parameters:
<ReferenceValue>
Default unit: DBM
Example:
CONF:CME:REF MAN
CONF:CME:REF:VAL -30
Manual operation:
See
CONFigure:CMEasurement:REFerence[:MODE]
<Mode>
Determines how the reference value for the compression point measurement is
defined.
Parameters:
<Mode>
AUTO | MANual
AUTO
The reference value is determined automatically.
MANual
The reference value is defined using
Configuring and Performing Measurements