![Ronde & Schwarz 1304.0004K03 User Manual Download Page 626](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987626.webp)
Remote Commands
R&S
®
FPL1000
626
User Manual 1178.3370.02 ─ 11
//--------------Performing the Measurement-----
//Select single sweep mode.
INIT:CONT OFF
//Initiate a new measurement and wait until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Query the results for the 3 dB and 5 dB compression points.
FETC:CME:P3DB?
//Result: -25,-22
FETC:CME:PNDB?
//Result: -22,-20
10.6.4
Measuring the Channel Power and ACLR
All remote control commands specific to channel power or ACLR measurements are
described here.
See also
Chapter 10.6.2, "Configuring Power Measurements"
Managing Measurement Configurations
............................................................... 626
......................................................................................628
..................................................................................... 631
Selecting the Reference Channel
......................................................................... 633
Performing an ACLR Measurement
......................................................................641
Retrieving and Analyzing Measurement Results
.................................................. 641
Programming Examples for Channel Power Measurements
................................ 642
10.6.4.1
Managing Measurement Configurations
The following commands control measurement configurations for ACLR measure-
ments.
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
............................................. 626
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
........................... 627
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete
.............................. 627
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE
.................................627
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
<Standard>
This command loads a measurement configuration.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Configuring and Performing Measurements