![Ronde & Schwarz 1304.0004K03 User Manual Download Page 460](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987460.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
460
User Manual 1178.3370.02 ─ 11
File contents
Description
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz for span > 0, s for span = 0, dBm/dB for statistics mea-
surements
Ref Level;-30;dBm
Reference level
Level Offset;0;dB
Level offset
Ref Position;75; %
Position of reference level referred to diagram limits (0 % =
lower edge)
y-Axis;LOG;
Scaling of y-axis linear (LIN) or logarithmic (LOG)
Level Range;100;dB
Display range in y direction. Unit: dB with x-axis LOG, % with x-
axis LIN
Rf Att;20;dB
Input attenuation
RBW;100000;Hz
Resolution bandwidth
VBW;30000;Hz
Video bandwidth
SWT;0.005;s
Sweep time
Trace Mode;AVERAGE;
Display mode of trace: CLR/WRITE,AVER-
AGE,MAXHOLD,MINHOLD
Detector;AUTOPEAK;
Selected detector
Sweep Count;20;
Number of sweeps set
Data section
Trace 1:;;
Selected trace
x-Unit;Hz;
Unit of x values: Hz with span > 0; s with span = 0; dBm/dB
with statistics measurements
y-Unit;dBm;
Unit of y values: dB*/V/A/W depending on the selected unit
with y-axis LOG or % with y-axis LIN
Values; 1001;
Number of measurement points
Frames;2;
Number of exported frames
Frame;0;
Most recent frame number
Timestamp;17.Mar 11;11:27:05.990
Timestamp of this frame
10000;-10.3;-15.7
10130;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values, identical to spectrum data:
<x value>, <y1>, <y2>; <y2> being available only with detector
AUTOPEAK and containing in this case the smallest of the two
measured values for a measurement point.
Frame;-1;
Next frame
Timestamp;17.Mar 11;11:27:05.342
Timestamp of this frame
...