![Ronde & Schwarz 1304.0004K03 User Manual Download Page 647](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987647.webp)
Remote Commands
R&S
®
FPL1000
647
User Manual 1178.3370.02 ─ 11
DISP:TRAC:Y:RLEV -10dBm
//Activate occupied bandwidth measurement.
CALC:MARK:FUNC:POW:SEL OBW
//Set the percentage of power to 99%.
POW:BWID 99PCT
//Set the channel bandwidth to 21 kHz.
POW:ACH:BAND 21kHz
//Optimize the instrument settings according to the channel bandwidth.
POW:ACH:PRES OBW
//Determine the ideal reference level for the measurement.
POW:ACH:PRES:RLEV
//Set the trace detector to positive peak.
DET APE
//--------------Performing the Measurement-----
//Select single sweep mode.INIT:CONT OFF
//Initiate a new measurement and waits until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Return the occupied bandwidth.
CALC:MARK:FUNC:POW:RES? OBW
10.6.7
Measuring the Spectrum Emission Mask
All remote control commands specific to spectrum emission mask measurements are
described here.
See also
Chapter 10.6.2, "Configuring Power Measurements"
Remote commands exclusive to spectrum emission mask measurements:
Managing Measurement Configurations
............................................................... 647
................................................................................649
Configuring a Multi-SEM Measurement
................................................................ 650
...................................................................................... 651
Configuring the Reference Range
.........................................................................665
.............................................................................667
Configuring the List Evaluation
............................................................................. 672
........................................................................673
.................................................................................................673
............................................................................... 674
10.6.7.1
Managing Measurement Configurations
The following commands control measurement configurations for SEM measurements.
Configuring and Performing Measurements