Remote Commands
R&S
®
FPL1000
632
User Manual 1178.3370.02 ─ 11
[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
.............................................................. 633
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
................................................ 633
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
................................................. 633
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
<Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
<Value>
This command defines the alpha value for the weighting filter for all channels.
Parameters:
<Value>
*RST:
0.22
Example:
POW:ACH:FILT:ALPH:ALL 0.35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
<Alpha>
This command defines the roll-off factor for the alternate channel weighting filter.
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
<Alpha>
This command defines the roll-off factor for the transmission channel weighting filter.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
Configuring and Performing Measurements