![Ronde & Schwarz 1304.0004K03 User Manual Download Page 757](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987757.webp)
Remote Commands
R&S
®
FPL1000
757
User Manual 1178.3370.02 ─ 11
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
//Activate signal tracking to keep the center frequency on the signal peak.
//After each sweep the maximum on trace 1 is searched within a range of 20 MHz
//around the center frequency. It must have a minimum power of -90dBm.
CALC:MARK:FUNC:STR ON
CALC:MARK:FUNC:STR:BAND 20MHz
CALC:MARK:FUNC:STR:THR -90dBm
CALC:MARK:FUNC:STR:TRAC 1
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BWIDth
<Bandwidth>
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth
<Bandwidth>
This command defines the bandwidth around the center frequency that is included in
the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Bandwidth>
Range:
10 Hz to Max span
*RST:
(= span/10 on activating the function)
Default unit: Hz
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
<Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Level>
The unit depends on
Range:
-130 dBm to 30 dBm
*RST:
-120 dBm
Default unit: DBM
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters