![Ronde & Schwarz 1304.0004K03 User Manual Download Page 831](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987831.webp)
Remote Commands
R&S
®
FPL1000
831
User Manual 1178.3370.02 ─ 11
CALC:SGR:TST:DATA? ALL
//Queries the time stamps of all stored frames.
CALC:SGR:FRAM:SEL -1
//Selects the frame that was captured 1 second after measurement start (Note the
//negative value!). This frame is displayed in the Spectrum window.
TRAC:DATA? SGR
//Retrieves the trace data for the spectrogram. For each frame, the power level
//and frequency at each sweep point are returned.
TRAC:DATA? TRACE1
//Retrieves the trace data for the selected frame only.
10.8.8
Working with Markers
The commands required to work with markers and marker functions in a remote envi-
ronment are described here. The tasks for manual operation are described in
In the Spectrum application, markers are identical in all windows. Thus, the suffix <n>
for the window is generally irrelevant.
............................................................................... 831
....................................................................................... 837
Configuring and Performing a Marker Search
.......................................................839
.......................................................................................... 843
.....................................................................................849
.............................................................................851
Fixed Reference Marker Settings
..........................................................................860
..................................................................................866
Phase Noise Measurement Marker
.......................................................................867
.............................................................................................. 870
................................................................................................ 874
............................................................................................. 877
............................................................................................879
Programming Examples for Using Markers and Marker Functions
10.8.8.1
Setting Up Individual Markers
The following commands define the position of markers in the diagram.
CALCulate<n>:DELTamarker<m>:AOFF
...........................................................................832
CALCulate<n>:DELTamarker<m>:LINK
............................................................................ 832
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
................................................832
CALCulate<n>:DELTamarker<m>:MODE
......................................................................... 833
CALCulate<n>:DELTamarker<m>:MREFerence
................................................................ 833
CALCulate<n>:DELTamarker<m>[:STATe]
........................................................................ 834
CALCulate<n>:DELTamarker<m>:TRACe
.........................................................................834
Setting Basic Spectrum RF Measurement Parameters