![Ronde & Schwarz 1304.0004K03 User Manual Download Page 905](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987905.webp)
Remote Commands
R&S
®
FPL1000
905
User Manual 1178.3370.02 ─ 11
Usage:
Query only
Manual operation:
See
CALCulate<n>:LIMit<li>:COPY
<Line>
This command copies a limit line.
Suffix:
<n>
.
<li>
Parameters:
<Line>
1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Manual operation:
See
CALCulate<n>:LIMit<li>:DELete
This command deletes a limit line.
Suffix:
<n>
.
<li>
Manual operation:
See
CALCulate<n>:LIMit<li>:STATe
<State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use
.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<State>
ON | OFF | 0 | 1
Setting Basic Spectrum RF Measurement Parameters