Getting Started
R&S
®
FPL1000
65
User Manual 1178.3370.02 ─ 11
Trace Information in Window Title Bar
Information on the displayed traces is indicated in the window title bar.
1
3
4
5
2
(1)
Trace color
Color of trace display in diagram
(2)
Trace no.
Trace number (1 to 6)
(3)
Detector
Selected detector:
AP
AUTOPEAK detector
Pk
MAX PEAK detector
Mi
MIN PEAK detector
Sa
SAMPLE detector
Av
AVERAGE detector
Rm
RMS detector
(4)
Trace Mode
Sweep mode:
Clrw
CLEAR/WRITE
Max
MAX HOLD
Min
MIN HOLD
Avg
AVERAGE (Lin/Log/Pwr)
View
VIEW
(5)
Smoothing factor
Smth
Smoothing factor, if enabled.
(See
4.4.1.3
Marker Information
Marker information is provided either in the diagram grid or in a separate marker table,
depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last two markers or delta
markers that were set are displayed, if available, as well as their index. The value in
the square brackets after the index indicates the trace to which the marker is assigned.
(Example: M2[1] defines marker 2 on trace 1.) For more than two markers, a separate
marker table is displayed beneath the diagram by default.
Operating the Instrument