Remote Commands
R&S
®
FPL1000
665
User Manual 1178.3370.02 ─ 11
[SENSe:]ESPectrum<sb>:SSETup
<State>
Enables or disables symmetrical configuration of the range settings.
See
Suffix:
<sb>
.
1 to 3
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Manual operation:
See
10.6.7.5
Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
...................................................................................... 665
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
.................................................................665
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe]
................................................................666
................................................................................666
.................................................................................... 666
[SENSe:]ESPectrum<sb>:BWID
<Bandwidth>
This command defines the channel bandwidth of the reference range.
The bandwidth is available if the power reference is the channel power.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Parameters:
<Bandwidth>
minimum span ≤ value ≤ span of reference range
*RST:
3.84 MHz
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
<Alpha>
This command defines the roll-off factor for the RRC filter.
The RRC filter is available if the power reference is the channel power.
Configuring and Performing Measurements