![Ronde & Schwarz 1304.0004K03 User Manual Download Page 815](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987815.webp)
Remote Commands
R&S
®
FPL1000
815
User Manual 1178.3370.02 ─ 11
Parameters:
<History>
The maximum number of frames depends on the number of
sweep points.
Range:
781 to 20000
Increment: 1
*RST:
3000
Example:
//Set the history depth to 1500
CALC:SGR:SPEC 1500
Manual operation:
See
CALCulate<n>:SGRam:LAYout
<State>
CALCulate<n>:SPECtrogram:LAYout
<State>
This command selects the state and size of spectrograms.
The command is available for result displays that support spectrograms.
Suffix:
<n>
.
Parameters:
<State>
ON
Spectrogram and trace diagram share a window.
OFF
Only the trace diagram is displayed, the spectrogram is not.
*RST:
OFF
Example:
CALC4:SPEC:LAY FULL
Shows the spectrogram in window 4. The corresponding trace
diagram is hidden.
Manual operation:
See
CALCulate<n>:SGRam[:STATe]
<State>
CALCulate<n>:SPECtrogram[:STATe]
<State>
This command turns the spectrogram on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:SGR ON
Activates the Spectrogram result display.
Setting Basic Spectrum RF Measurement Parameters