Remote Commands
R&S
®
FPL1000
719
User Manual 1178.3370.02 ─ 11
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results.
//If the results are not accurate, change the position of the
//the temporary markers manually.
//----Changing the position of the temp markers-----
CALC:MARK:X 100MHZ
//Positions the reference marker on 100 MHz.
CALC:DELT2:X 5KHZ
//Positions delta marker 2 and 3 at a distance of 5 kHz to the reference marker.
CALC:DELT3:X 1KHZ
//Corrects the position of delta marker 3 by 1 kHz.
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results for the repositioned markers.
10.6.14
Remote Commands for EMI Measurements
The following commands are required to perform EMI measurements in a remote envi-
ronment. This measurement requires the R&S
FPL1-K54 option.
The following tasks specific to the EMI application are described here:
................................................................................ 719
...................................................................................... 720
Configuring the EMI Final Test
..............................................................................721
.................................................................................. 722
..........................................................................................724
.......................................................................................... 725
Programming Example: EMI Measurement
.......................................................... 725
10.6.14.1
Activating EMI Measurement
EMI measurement must be activated explicitely.
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement[:STATe]
.................................. 719
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement[:STATe]
......................................... 719
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement[:STATe]
<State>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement[:STATe]
<State>
This command turns the EMI measurement marker functionality on and off.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
Configuring and Performing Measurements